Your Source for Process Measurement & Control
UK
44-796-072-6366
Germany
49-178-631-1266
Product Groups
Multifunction Process Calibrator with RS232 Interface- High Accuracy
TRODEKS Engineering | Thermocouples, Pressure Transducers, Flow Meters, PID Controllers
PCL1000
PLACE ORDER 
  • Excellent Accuracy ±0.6°C T/C, ±0.3°C RTD 0.015% Electrical 0.05% Pressure
  • Read and Source Modes
  • 10 T/C Types; 8 RTD types plus 400 Series Thermistor
  • Stores up to 9 setpoints for each output function
  • Measures pressure using external modules
 Manuals (contact with email : [email protected])
 Print
Description

Description

SPECIFICATIONS
Operational Temp Range:
0 to 50°C (14 to 122°F)
Stability:0.01% F.S./°C outside of 23°C±5°C
Pressure Accuracy: 0.05% FS
Max Overpressure: 300%
Ports:<100 psi 1/4" barbed hose fittings,≤ 100 psi: 1/8" brass swagelok fittings
Wetted Parts: clean dry gases only
T/C Types: J,K,T,E,R,S,B,N,L,U
RTD Ranges: RTD Ranges: Pt385 (100, 200, 500, 1000&trodeks;) Pt392, JIS, Ni120, CU10, YS1400
Ohm Ranges: 0 to 400.00 and 400.0 to 4000.0
Voltage:
(Low) I/O 0-20.000 V
(High) Input only 0-250 V ac/dc
Frequency: 0-1000.0 CPS, Hz 0-10.00 kHz
Continuity: 0 to 400.0&trodeks; audible tone under 10&trodeks;
Accuracy: T/C Type J: ±0.3°C ± 10µV (±0.6°C total error)
RTD Pt100: ±0.1°C ±0.075 &trodeks; (±0.3°C total error)
Ohms (400): ±0.1 &trodeks;
Ohms (4000): ±1.0 &trodeks;
mA Compliance: up to 1000 &trodeks;
Voltage Drive: 1 mA
RTD IEX-range: 0.01 mA to 3 mA
RTD Frequency Response: 10 mS
Operating Temp: -10 to 50°C
Storage Temp: -40 to 60°C
Power: 4 AA alkaline batteries (included)
Dimensions: 191x102x38 mm(7.5x4x1.5")
Weight: 350g (12oz)

External Pressure Modules*
MODEL NO. Parameter Range Gauge (psig) Accuracy Over Pressure
PCL-PMA Pressure Module adaptor (one required)
PCL-PM030G 0 to 30 (0 to 2 bar) ±0.025% 300%
PCL-PM050G 0 to 50 (0 to 3.5 bar) ±0.03% 300%
PCL-PM100G 0 to 100 (0 to 7 bar) ±0.025% 300%
PCL-PM300G 0 to 300 (0 to 20 bar) ±0.025% 200%
PCL-PM1KG 0 to 1000 (0 to 70 bar) ±0.05% 200%
PCL-PM3KG 0 to 3000 (0 to 200 bar) ±0.1% 200%
PCL-PM5KG 0 to 5000 (0 to 340 bar) ±0.1% 200%
PCL-PM015VAC 0 to -15 (0 to -1 mbar) ±0.025%, ±0.0025 psi 300%
PCL-PM015A 0 to 15 (0 to 1 bar) abs. ±0.025%, ±0.0025 psi 300%
PCL-PM030A 0 to 30 (0 to 2 bar) abs. ±0.025% 300%
PCL-PM015C -15 to 15 (-1 to 1 bar) ±0.025%, ±0.0025 psi 300%
PCL-PM030C -15 to 30 (-1 to 2 bar) ±0.025%, ±0.0025 psi 300%
PCL-PM030D 0 to 30 (0 to 2 bar) diff. ±0.025% 300%
* Most popluar ranges (see pressure module page for all ranges).
Anemometer  |   Automation  |   Accelerometers  |   Calibrator  |   Cartridge Heaters  |   Chart Recorders  |   Cryogenics  |   Custom Pressure Transducers  |   Data Loggers  |   Data Acquisition  |   Dial Thermometers  |   Fiber Optics  |   Fiber Optic Temperature Measurement  |   Flexible Heaters  |   Flow Meters  |   Glass Bulb Thermometers  |   Grippers  |   Heaters  |   Humidity Measurement  |   Integrated Circuit Sensors  |   Immersion Heaters  |   Infrared Thermometers  |   Level Measurement  |   Load Cells  |   M12 Temperature Connectors  |   Magnetic Flow Meter  |   Panel Meter  |   PID Controllers  |   pH meter  |   Pressure Gauges  |   Pressure Transducers  |   Pressure Transmitters  |   Rotameters  |   RTD  |   Remote Process Monitoring  |   Sanitary Sensors  |   Signal Conditioners  |   Stepper Motors  |   Strain Gages   |   Strip Heaters  |   Temperature Controllers  |   Temperature Labels  |   Temperature Measurement  |   Temperature Sensor Manufacturing  |   Temperature Sensors  |   Temperature Transmitters  |   Thermal Imagers  |   Thermistors  |   Thermocouple  |   Thermocouple Wire | Thermowell  |   Ultrasonic Flow Meters  |   Vortex Flow Meter  |   Wireless Sensors
trodeks.comtrodeks.com - فقخیثنس.زخئ - صصص.فقخیثنس.زخئ