High Flowrate Mass Controllers and Meters for All Clean Gases
FMA-700A and FMA-800A
PLACE ORDER - ±1% Accuracy (Up to 500 SLM), ±0.2% Repeatability
- Low Cost
- For Flowrates Up to 1000 SLM
- Normally Closed Valve for Added Safety
- Fast Response
- Altitude Insensitive
- Removable Sensing Tube Assembly
Description
Description
The trodeks® FMA-700A and 800A Series mass flowmeters and controllers extend the range of mass flow measurement and control from 0 to 5 SCCM to 1000 SLM with 50:1 turndown capability. Like all mass flowsensors, these new units measure mass flow rate directly, without the need for temperature or pressure corrections.
The FMA-700A and 800A Series work on the heated bypass sensor tube principle. The units feature corrosion resistant stainless steel and FKM construction and a 325 mesh (44 micron) inlet screen to preventcontamination of the sensor. The flow body and valve leak less than 10-8 atm-cc/sec of helium. In the event of a power failure, the valve is normally closed.
The FMA-78P2 and FMA-78P4 are 2- and 4-channel readout/power supply/setpoint control electronicboxes for use with FMA-700A and 800A Series mass flow sensors. The units feature 000.0 to 102.4%of full scale display, a ±15 Vdc power supply, setpoint potentiometers (for mass flow controllers), power cord, and an electrical connector for hookup of the electronics boxes’ outputs.
The FMA-78P Series rate indicators/electronics packages also have back-panel blend control switches that select independent/slave operations for each channel (channel 1 is the master), and the units include a 0 to 5 Vdc output for each channel for chart recording, data acquisition, or computer interfacing.
SPECIFICATIONS
Accuracy: ±1% full scale (up to 500 SLM), ±2% full scale for 750 to 1000 SLM
Repeatability: ±0.2% full scale
Turndown Ratio: 50:1
Response Time: 2.0 seconds
Analog Setpoint:
0 to 5 Vdc: Voltage output models
4 to 20 mA: Current output models
Linear Output:
Standard: 0 to 5 Vdc into 2000 &trodeks; min.
Optional: 4 to 20 mA, into 200 to 750 &trodeks;
Max Pressure:
FMA-760A to FMA-772A: 1000 psig
FMA-773A to 778A: 200 psig
FMA-860A to FMA-874A: 1500 psig
FMA-875A to FMA-878A: 1000 psig
Pressure Coefficient: 0.1% typical
Differential Pressure:
FMA-800A Series: 0.5 psid
FMA-700A Series: No less than 20% of supply pressure, depending on parameters
Temperature Range (Gas and Ambient): -10 to 70°C (14 to 158°F)
Temperature Coefficient: 0.05% full scale/°C
Wetted Parts: 316 SS body and sensing tube assembly, FKM elastomers, other materials available
POWER
Voltage Output Models: 12 Vdc (±5%) or 15 Vdc (±10%)
mA Output Models: 15 Vdc (±5%) or 24 Vdc (±15%)
Current Consumption:
FMA-800A: <45 mA DC
FMA-700A: <250 mA DC
Anemometer | Automation | Accelerometers | Calibrator | Cartridge Heaters | Chart Recorders | Cryogenics | Custom Pressure Transducers | Data Loggers | Data Acquisition | Dial Thermometers | Fiber Optics | Fiber Optic Temperature Measurement | Flexible Heaters | Flow Meters | Glass Bulb Thermometers | Grippers | Heaters | Humidity Measurement | Integrated Circuit Sensors | Immersion Heaters | Infrared Thermometers | Level Measurement | Load Cells | M12 Temperature Connectors | Magnetic Flow Meter | Panel Meter | PID Controllers | pH meter | Pressure Gauges | Pressure Transducers | Pressure Transmitters | Rotameters | RTD | Remote Process Monitoring | Sanitary Sensors | Signal Conditioners | Stepper Motors | Strain Gages | Strip Heaters | Temperature Controllers | Temperature Labels | Temperature Measurement | Temperature Sensor Manufacturing | Temperature Sensors | Temperature Transmitters | Thermal Imagers | Thermistors | Thermocouple | Thermocouple Wire | Thermowell | Ultrasonic Flow Meters | Vortex Flow Meter | Wireless Sensors