Mass Flow Device
FMA-7100, FMA-7200 and FMA-7300
PLACE ORDER - ±1% Full Scale Accuracy
- ±0.25% of Rate Repeatability
- Broad Range in a Single Unit to Fit Most Applications
- Field Selectable Soft Start Eliminates Spoiled Batches
- Normally Closed Valve
- Low Pressure Drop
- Compact Design
- For Clean, Non-Coating Gases
Description
Description
trodeks® FMA-7000E Mass Flow Controllers accurately measure gases on the basis of mass flow. The flow sensor produces an electrical output signal linear with the flowrate used for indicating, recording, and/or control purposes. It eliminates the need for continuous monitoring and readjustment of gas pressures to provide a stable gas flow. This is accomplished automatically by an electromagnetically actuated flowrate control valve.
OPERATING PRINCIPLE The operating principle of the controller is thermodynamic. A precision power supply directs heat to the midpoint of the sensor tube carrying the flow. On the same tube, equidistant upstream and downstream of the heat input, are resistance temperature measuring elements. With no flow, the heat reaching each temperature element is equal. With increasing flow, the flow stream carries heat away from the upstream element, and carries it towards the downstream element. An increasing temperature difference develops between the two elements and this difference is proportional to the mass flowrate of the gas. A bridge circuit interprets the temperature difference and an amplifier provides the output to the control circuitry as well as a 0-5 Vdc output signal.
SPECIFICATIONS
Accuracy: ±1% FS including linearity at calibration conditions, ±1.5% FS including linearity for flow ranges greater than 20 SLM
Repeatability: ±0.25% of rate
Response Time: <3 second response to within 2% FS final value with a 0 to 100% command step
Set Point Input: 0 to 5 Vdc or 1000Ω potentiometer
Output: 0 to 5 Vdc into 2000Ω minimum load; 3 mV rms max. ripple
Maximum Pressure: 1500 psig, pressure drop 5 to 50 psid
Temperature Range: 4 to 66°C(40 to 150°F) ambient and gas
Temp. Sensitivity:
Zero: < ±0.075% FS per °C
Span: <±1.0% FS shift over 10 to 50°C (50 to 122°F) range
Power Supply Sensitivity: ±0.09% FS per % power supply voltage variation Mounting Attitude Sensitivity: ±0.5% maximum FS deviation after zeroing
Leak Integrity: 1 x 10-9 Atm. scc/sec helium
Control Range: 50 to 1
Pressure Sensitivity: ±0.03% per psi up to 200 psig (N2)
Power: +15 Vdc at 35 mA DC, -15 Vdc at 180 mA dc, 3.5 watts power consumption
Wetted Parts: 316 and 430SS, FKM
Connections: ¼" stainless steel compression fittings
Electrical Connections: Card edge, 20 terminals, gold over low stress nickel-plated copper
Accessories Included: Mating connector with 1.5 m (5') ribbon cable model. Easy connector/cable assemblies available
Anemometer | Automation | Accelerometers | Calibrator | Cartridge Heaters | Chart Recorders | Cryogenics | Custom Pressure Transducers | Data Loggers | Data Acquisition | Dial Thermometers | Fiber Optics | Fiber Optic Temperature Measurement | Flexible Heaters | Flow Meters | Glass Bulb Thermometers | Grippers | Heaters | Humidity Measurement | Integrated Circuit Sensors | Immersion Heaters | Infrared Thermometers | Level Measurement | Load Cells | M12 Temperature Connectors | Magnetic Flow Meter | Panel Meter | PID Controllers | pH meter | Pressure Gauges | Pressure Transducers | Pressure Transmitters | Rotameters | RTD | Remote Process Monitoring | Sanitary Sensors | Signal Conditioners | Stepper Motors | Strain Gages | Strip Heaters | Temperature Controllers | Temperature Labels | Temperature Measurement | Temperature Sensor Manufacturing | Temperature Sensors | Temperature Transmitters | Thermal Imagers | Thermistors | Thermocouple | Thermocouple Wire | Thermowell | Ultrasonic Flow Meters | Vortex Flow Meter | Wireless Sensors