The CDCN441 conductance/resistance analyzercontrolleris designed for online monitoring of processsolutions and water applications. Applications include:waste water treatment, power plants, thermo electric plants,pharmaceutical, photographic industries and soft drinksindustries. The menus are self guiding with simple three buttonprogramming. Available control outputs include ON-OFF orpulse width modulation. The transmitter output is alsoprogrammable from the keypad eliminating the need forpotentiometer or dip-switch settings. The calibration functionautomatically recognizes the cell used and the buffer solutionresulting in a simple and straight forward system calibration.
Equipment is built with solid state technology, electroniccontacts and not electrical contacts, avoiding mechanicalmovement and off course no sparks.
SPECIFICATIONS
General
Construction/Materials:
Case: Aluminum (SAE323)
Faceplate: ABS
Anti-Corrosion Treatment: Finished with electrostatic epoxy paint
Case Rating: NEMA 4 (IP68)
Power Consumption: 3.5 VA
Operation: 90-240 Vac; 50/60Hz
LCD Readout: 2 lines x 16 characters
Reading Modes: Continuous, average or hold
Assembly in 51 mm (2") Tube or Flat Surface or Panel: 144 x 144 x 100 mm DIN (5.7 x 5.7 x 4")
Weight: 1.3 kg (2.8 lb)
RS485 Output: MODBUS and proprietary communication protocol
Analyzer
Conductivity Range: 1µS/cm to 2S/cm
Resistivity Range: 10M&trodeks;/cm to infinity
Resolution: 0.1 or 0.01
Temperature Compensation: 0 to 100°C (32 to 212°F), Automatic or Manual
Operating Temperature: 5 to 40°C (41 to 104°F)
Cell Constants Offered: 0.01, 0.1, 1, 5 or 10 cm-1
Transmitter
Analog Output: 4 to 20 mA (2), this output can be programmed for transmission
Impedance: 600 &trodeks;
Optical Galvanic Isolation: 2000 Vac
Controller Outputs: 2 N.O. (1A/240 Vac) on/off alarm or PWM; 4 to 20 mA for PID control or retransmittion
Cells
CDE-440-001:
Range: 0.01 µS to 2 mS in SS or Titanium, K = 0.01, 0 to 100°C (32 to 212°F), 10 kg/cm2
Insertion Length: 54 mm (2.1")
Thread: 316 SS, ¾NPT
CDE-440-01:
Range: 0.1 µS to 20 mS, K = 0.1, 0 to 130°C (32 to 266°F), 10 kg/cm2
Insertion Length: 42 mm (1.7")
Thread: 316 SS, ¾NPT
CDE-440-1:
Range: 0.1µS to 100 mS, K = 1, 0 to 130°C (32 to 266°F), 10 kg/cm2
Insertion Length: 60 mm (2.4")
Thread: 316 SS, ¾NPT
CDE-440-5:
Range: 1 µS to 150 mS, K = 5, 0 to 130°C (32 to 266°F), 10 kg/cm2
Insertion Length: 60 mm (2.4")
Thread: 316 SS, ¾NPT (PVDF body)
CDE-440-01T:
Range: 0.1 µS to 20 mS, K = 0.1, 0 to 200°C (32 to 392°F), 10 kg/cm2
Insertion Length: 90 mm (3.5")
Thread: 316 SS, ¾NPT Anemometer | Automation | Accelerometers | Calibrator | Cartridge Heaters | Chart Recorders | Cryogenics | Custom Pressure Transducers | Data Loggers | Data Acquisition | Dial Thermometers | Fiber Optics | Fiber Optic Temperature Measurement | Flexible Heaters | Flow Meters | Glass Bulb Thermometers | Grippers | Heaters | Humidity Measurement | Integrated Circuit Sensors | Immersion Heaters | Infrared Thermometers | Level Measurement | Load Cells | M12 Temperature Connectors | Magnetic Flow Meter | Panel Meter | PID Controllers | pH meter | Pressure Gauges | Pressure Transducers | Pressure Transmitters | Rotameters | RTD | Remote Process Monitoring | Sanitary Sensors | Signal Conditioners | Stepper Motors | Strain Gages | Strip Heaters | Temperature Controllers | Temperature Labels | Temperature Measurement | Temperature Sensor Manufacturing | Temperature Sensors | Temperature Transmitters | Thermal Imagers | Thermistors | Thermocouple | Thermocouple Wire | Thermowell | Ultrasonic Flow Meters | Vortex Flow Meter | Wireless Sensors