Your Source for Process Measurement & Control
UK
44-796-072-6366
Germany
49-178-631-1266
Product Groups
3-1-2 Digit Thermocouple Meters- 1-8 DIN
TRODEKS Engineering | Thermocouples, Pressure Transducers, Flow Meters, PID Controllers
DP302-J, -K, -T
PLACE ORDER 
  • J, K or T Thermocouple Input
  • 1° or 0.1° Resolution
  • °C or °F, Jumper- Selectable
  • 0.05 Deg/Deg Cold- Junction Error
  • Five-Segment Linearization
  • 1 mV/deg Linearized Recorder Output
  • 120 dB CMR, 70 dB NMR
  • Bright, 14.2 mm (0.56") LED Display
  • Display Hold and Test
  • 115/230 Vac Power
  • EMI/RFI Filter for AC Power
  • Screw-Terminal Barrier Strip
  • Short 104 mm (4.1") Deep 1/8 DIN Case Options
  • Isolated 9 to 32 Vdc Power or 26 to 56 Vdc Power
  • NEMA 4 (IP65) Splash-Proof Lens Cover
 Manuals (contact with email : [email protected])
 Print
Description

Description

The DP302-J, -K and -T are 3-1/2 digit (±1999 count)panel meters for thermocouple types J, K and T.They are available for display in °C or °F andresolution of 1° up to 1999° or 0.1° up to 199.9°. Theyare recommended for demanding indication-onlyapplications where exceptionally low-cost is required.For display up to 1999.9° with 0.1° resolution,trodeks®. recommends its DP3002 Seriesmeters, which provides a 4-1/2 digit (±19,999 count)main assembly.

STANDARD FEATURES
The DP302 Series thermocouple meters provideexceptional accuracy, as made possible by fivesegmentlinearization and use of matched, gradedcomponents for each T/C type. Electrical featuresinclude cold-junction compensation and jumperselection of °C or °F. A 1 mV per count linearizedanalog output is standard and is ideal for driving astrip-chart recorder.

Mechanical features include a bright 14.2 mm (0.56")display, a screw-terminal barrier strip for signal andpower, and a compact 1/8 DIN case that requires less

Electrical features include cold-junction compensation,jumper selection of °C or °F, and a linearized recorderoutput of 1 mV per degree.

OPTIONS
Options are isolated 9 to 32 Vdc or 26 to 56 Vdcpower, a PCB edge connector for display control andoutput of 4.7 Vdc and -4.7 Vdc, and a splash-prooflens cover which meets NEMA 4 (IP65) requirements.

SPECIFICATIONS
Input
Thermocouple:
  Type J: (Iron-Constantan, Fe-CuNi)
  Type K: (Chromel-Alumel, NiCr-NiAl)
  Type T: (Copper-Constantan, Cu-CuNi)
Calibration: IEC 584-1 (IPTS-68)
Configuration: Single-ended (-TC lead connected to ANA GND)
Polarity: Bipolar
Zero: Adjustable ±20°C ( ±36°F)
Cold-Junction Tempco: ±0.05 deg/deg
Sensor-Wire Resistance Effect Per Conductor:
  Type J: 40 µdeg/deg/Ω, up to 2500 Ω
  Type K: 50 µdeg/deg/O, up to 2000 Ω
  Type T: 50 µdeg/deg/O, up to 2000 Ω
Sensor-Break Current: 300 nA
Sensor-Break Indication: Meter displays positive overrange (upscale)
Noise Rejection
NMR, SIG HI to SIG LO: 70 dB, 50/60 Hz
CMR, ANA GND to PWR GND: 120 dB, DC to 60 Hz
CMV, ANA GND to PWR GND: 1500 Vp per HV test, 354 Vp per IEC spacing
Digital Inputs
Level: TTL or 5 V CMOS compatible
Accuracy at 25°C (1° resolution models)
Range for Rated Accuracy:
  Type J: - 138 to 760°C (- 216 to 1400°F)
  Type K: - 105 to 1248°C (- 157 to 1999°F)
  Type T: - 112 to 400°C (- 170 to 752°F)
Maximum Error: ±1.5°C (±2.7°F) ±1/2 count
Span Tempco: ±0.01% of reading/°C
Resolution: 0.1°C or 0.1°F (jumper-selectable)
Full-Scale Step Response: 1 s
Warmup to Rated Accuracy: 10 min
Accuracy at 25°C (0.1° resolution models)
Range for Rated Accuracy, all T/C Types: - 105.0 to 199.9°C (- 157.0 to 199.9°F)
Maximum Error: ±1.0°C (±1.8°F) ±1/2count
Span Tempco: ±0.01% of reading/°C
Resolution: 0.1°C or 0.1°F (jumper-selectable)
Full-Scale Step Response: 1 s
Warmup to Rated Accuracy: 20 min
Analog Output (Linearized)
Voltage: 1 mV/count
Calibration Error: ±1.5 mV on °C, ±2.7 mV on °F
Current: 1 mA max
Source Resistance: 68 Ω
Analog-to-Digital Conversion
Technique: Dual-slope, average value
Signal-Integration Period: 100 ms
Read Rate: 2.5/s
Display
Type: 7-segment, red LED
Height: 14.2 mm (0.56")
Symbols: - 1.8.8.8
Overrange Indication: Three leastsignificant digits blank
Power
AC Voltages: 115 or 230 Vac, ±15%
AC Frequency: 49 to 440 Hz
DC Voltages: 9 to 32 Vdc, isolated to 300 Vp; 26 to 56 Vdc, isolated to 300 Vp
Power Consumption: 3.7 W
Output Voltages: 4.7 Vdc and - 4.7 Vdc ±5%, 10 mA max
Environmental
Operating Temperature: 0 to 60°C
Storage Temperature: - 40 to 85°C
Relative Humidity: 95% at 40°C (non-condensing)
Mechanical
Bezel: 96 W x 48 H x 5.1 mm D (3.78 x 1.89 x 0.20")
Depth Behind Bezel: 104 mm (4.09")
Panel Cutout: 92 W x 45 mm H (3.62 x 1.77")
Weight: 15 oz (425 g)
Case Material: 94V-0 UL-rated polycarbonate
Anemometer  |   Automation  |   Accelerometers  |   Calibrator  |   Cartridge Heaters  |   Chart Recorders  |   Cryogenics  |   Custom Pressure Transducers  |   Data Loggers  |   Data Acquisition  |   Dial Thermometers  |   Fiber Optics  |   Fiber Optic Temperature Measurement  |   Flexible Heaters  |   Flow Meters  |   Glass Bulb Thermometers  |   Grippers  |   Heaters  |   Humidity Measurement  |   Integrated Circuit Sensors  |   Immersion Heaters  |   Infrared Thermometers  |   Level Measurement  |   Load Cells  |   M12 Temperature Connectors  |   Magnetic Flow Meter  |   Panel Meter  |   PID Controllers  |   pH meter  |   Pressure Gauges  |   Pressure Transducers  |   Pressure Transmitters  |   Rotameters  |   RTD  |   Remote Process Monitoring  |   Sanitary Sensors  |   Signal Conditioners  |   Stepper Motors  |   Strain Gages   |   Strip Heaters  |   Temperature Controllers  |   Temperature Labels  |   Temperature Measurement  |   Temperature Sensor Manufacturing  |   Temperature Sensors  |   Temperature Transmitters  |   Thermal Imagers  |   Thermistors  |   Thermocouple  |   Thermocouple Wire | Thermowell  |   Ultrasonic Flow Meters  |   Vortex Flow Meter  |   Wireless Sensors
trodeks.comtrodeks.com - فقخیثنس.زخئ - صصص.فقخیثنس.زخئ