Your Source for Process Measurement & Control
UK
44-796-072-6366
Germany
49-178-631-1266
Portable- High Accuracy Multifunction Calibrator
PCL1200
PLACE ORDER - Measures and Sources T/C (13 Types), RTDs (13 Types), Ohms, Current, Voltage, Frequency
- Sources Pulse Trains
- Isolated mA/V Readback Circuit for Complete Transmitter Calibration
- Pressure Module Communication Port Compatible with trodeks
PCL-PM Series Pressure Modules - Built-In 24 V Supply Can Drive 4 to 20 mA Loops Up to 1000 Ohms
- Direct Entry of Custom RTD Coefficients (R0, A, B, C)
- All Source Modes Can be Programmed with Dedicated Setpoints to Speed Calibration and Linearity Tests
- Highest Accuracy in Class—to 0.015% of Reading
- Meets CE Requirements and is Designed to IEC 1010 Safety Standards
- Supplied with Full Rubber Boot
Description
Description
The PCL1200 multifunctioncalibrator has a feature setunmatched by other high-accuracyhandheld calibrators in its pricerange. It is portable, provides thefunctions and accuracy of alaboratory-grade instrument, andhas everything needed for virtuallyany calibration task. It can alsomeasure and source thermocouples,RTDs, current, voltage, andfrequency, and it can source pulsetrains. It has a communications portfor pressure modules and anisolated mA/V readback circuit.Arrow keys, direct numeric keypad entry, and 3 software-drivenfunction buttons combine with alarge, backlit, menu-driven graphicsdisplay to provide a highly intuitiveyet powerful interface. A built-in250 Ohm resistor for HartTMcompatibility, compatibility withsmart transmitters and PLCs, fullfuseless protection, and a serialcommunications port for controlusing ASCII commands areadditional features that make thePCL1200 indispensable forcalibration tasks. A protective rubber boot is included with each unit.
Measuring Pressurewith the PCL1200With a PCL-PMA pressure moduleadaptor, the PCL1200 will work withall trodeks PCL-PM pressuremodules. See Specs and"To Order" chart.
SPECIFICATIONS (23 ±5°C Unless Otherwise Noted)
Voltage Read and Source:
Source: 0 to 20 Vdc
Read:
Isolated: 0 to 30 Vdc
Non-Isolated: 0 to 20 Vdc
Thermocouple mV:
Read/Source: -10 to 75 mV
Current (mA):
Source: 0 to 24 mA
Read:
Isolated: 0 to 24 mA
Non-Isolated: 0 to 24 mA
Frequency (1 to 20 V SelectableAmplitude):
cpm Source/Read: 2 to 600 cpm
Hz Source/Read: 1 to 1000 Hz
kHz Source/Read: 1 to 10 kHz
Pulse (Source Only; 1 to 20 VSelectable Amplitude): 1 to 30,000pulses; 2 cpm to 10 kHz
Ohms:
Source: 5 to 4000 O
Read: 0 to 4000 O
Thermocouple Read and Source:
Type J: -200 to 1200°C(-328 to 2192ºF)
Type K: -200 to 1370°C(-328 to 2498ºF)
Type T: -200 to 400°C(-328 to 752ºF)
Type E: -200 to 950°C(-328 to 1742ºF)
Type R: -20 to 1750°C(-4 to 3182ºF)
Type S: -20 to 1750°C(-4 to 3182ºF)
Type B: 600 to 1800°C(1112 to 3272ºF)
Type C: 0 to 2316°C(32 to 4201ºF)
Type XK: -200 to 800°C(-328 to 1472ºF)
Type BP: 0 to 2500°C(32 to 4532ºF)
Type L: -200 to 900°C(-328 to 1652ºF)
Type U: -200 to 400°C(-328 to 752ºF)
Type N: -200 to 1300°C(-328 to 2372ºF)
RTD Read and Source:
Ni120 (672): -80 to 260°C(-112 to 500ºF)
Pt100 (385): -200 to 800°C(-328 to 1472ºF)
Pt100 (3926): -200 to 630°C(-328 to 1166ºF)
Pt100 (3916): -200 to 630°C(-328 to 1166ºF)
Pt200 (385): -200 to 630°C(-328 to 1166ºF)
Pt500 (385): -200 to 630°C(-328 to 1166ºF)
Pt1000 (385): -200 to 630°C(-328 to 1166ºF)
Cu10: -100 to 260°C(-148 to 500ºF)
YSI400: 15 to 50°C(59 to 122ºF)
Cu50: -180 to 200°C(-292 to 392ºF)
Cu100: -180 to 200°C(-292 to 392ºF)
Pt385-10: -200 to 800°C(-328 to 1472ºF)
Pt385-50: -200 to 800°C(-328 to 1472ºF)
Environmental
Operating Temperature: -10 to 50°C(14 to 122ºF)
Storage Temperature: -20 to 70°C(-4 to 158ºF)
Stability: ±0.005% of reading/°C(±5°C outside of 23°C)
Power Requirement: 6 Vdc
Batteries: 4 "AA" alkaline (included)or optional rechargeable
Mechanical
Dimensions:220.9 H x 106.6 W x 58.4 mm D(8.7 x 4.2 x 2.3")
Weight: 863 g (30.5 oz)
Optional Accessory: Carrying case(Model No. CC572)
AccuracyVoltage: ±0.015% of reading, ±2 mV
Thermocouple: mV ±0.02% of reading,±10 µV
Thermocouple Errors (In °C; Add 0.2for Cold-Junction CompensationError)
RTD Read/Source:
Type J: 0.2°C Ni120 (672) 0.2°C
Type K: 0.3°C Pt100 (385) 0.2°C
Type T: 0.2°C Pt100 (3926) 0.2°C
Type E: 0.2°C Pt100 (3916) 0.2°C
Type R: 1.2°C Pt200 (385) 0.8°C
Type S: 1.2°C Pt500 (385) 0.4°C
Type B: 1.2°C Pt1000 (385) 0.2°C
Type C: 0.6°C Cu10 1.4°C
Type XK: 0.2°C YSI400 0.1°C
Type BP: 0.9°C Cu50 0.4°C
Type L: 0.2°C Cu100 0.3°C
Type U: 0.25°C Pt385-10 1.4°C
Type N: 0.4°C Pt385-50 0.4°C
Read Source:
Current (mA): ±0.015% of reading,±2 µA ±0.015% of reading, ±2 µA
cpm: ±0.05% of reading, ±1 LSD±0.05% of reading
Hz: ±0.05% of reading, ±1 LSD±0.05% of reading
kHz: ±0.05% of reading, ±1 LSD±0.125% of reading
Ohms (Works with All PulsedTransmitters Down to 5 ms):
400 Ω Range: ±0.025% of reading,±0.05 Ω ±0.025% of reading,±0.05 Ω
4000 Ω Range: ±0.025% of reading,±0.5 Ω ±0.025% of reading, ±0.5 Ω Anemometer | Automation | Accelerometers | Calibrator | Cartridge Heaters | Chart Recorders | Cryogenics | Custom Pressure Transducers | Data Loggers | Data Acquisition | Dial Thermometers | Fiber Optics | Fiber Optic Temperature Measurement | Flexible Heaters | Flow Meters | Glass Bulb Thermometers | Grippers | Heaters | Humidity Measurement | Integrated Circuit Sensors | Immersion Heaters | Infrared Thermometers | Level Measurement | Load Cells | M12 Temperature Connectors | Magnetic Flow Meter | Panel Meter | PID Controllers | pH meter | Pressure Gauges | Pressure Transducers | Pressure Transmitters | Rotameters | RTD | Remote Process Monitoring | Sanitary Sensors | Signal Conditioners | Stepper Motors | Strain Gages | Strip Heaters | Temperature Controllers | Temperature Labels | Temperature Measurement | Temperature Sensor Manufacturing | Temperature Sensors | Temperature Transmitters | Thermal Imagers | Thermistors | Thermocouple | Thermocouple Wire | Thermowell | Ultrasonic Flow Meters | Vortex Flow Meter | Wireless Sensors
trodeks.comtrodeks.com - فقخیثنس.زخئ - صصص.فقخیثنس.زخئ