Your Source for Process Measurement & Control
UK
44-796-072-6366
Germany
49-178-631-1266
PORTABLE MASS FLOWMETER AND CALIBRATION KIT
FMA-PC16
PLACE ORDER - Ranges of 0 to 0.5 SCCM up to 0 to 1500 SLM
- <10 ms Response Time—Field Adjustable
- 30+ Gas Calibrations, Including Air, Ar, CH4, CO, CO2, Ethane, H2, He, N2, N2O, Neon, O2, Propane, Butane, Iso-Butane, Acetylene, Ethylene, Krypton, Xenon, and Sulfur Hexafluoride
- Pressure, Temperature and Volumetric and Mass Flow Simultaneously Displayed
- NIST 5-Point Certificate Included
- No Straight Runs of Pipe Required
- RS232 and USB Standard
Description
Description
trodeks’s FMA-PC16 Seriesportable digital mass flowmetersaccurately and rapidly measure themass flow rate, volumetric flow rate,pressure and temperature ofprocess gases.
The FMA-PC16 couples theFMA-1600A Series meter withfittings and tubing to create aportable gas chromatographycalibrator.
trodeks’s FMA-PC16 Seriesportable calibration units combineup to three FMA-1600A Series orFMA-LP1600A Series massflowmeters with totalizers andhigh-accuracy calibration inside arugged industrial carrying case withits own power supply and battery.
FMA-PC16 units include integratedpush-connect fittings for each meterand flow software. At the heart oftrodeks’s mass flowmeters andcalibrators is an internallycompensated, laminar differentialpressure measurement technologythat yields linear results over avery wide flow range. This producesvolumetric and mass flow data thatis fully compensated for changes inpressure and temperature andcorrected for the user’s STP.All portable mass flowmeters andcalibrators come with NIST-traceablecalibration sheets and excellentcustomer support, includingtechnical phone support.
trodeks’s FMA-1600A Series andFMA-PC16 Series mass flowmetersfeature large integrated displays withtouchpad accessible gas selectionfrom 30 user-selectable gascalibrations and both analog andRS232 digital communications.
Specifications
Accuracy: Calibration conditions after tare ±(0.8% of reading ± 0.2% of full scale)
High Accuracy: Calibration conditions after tare ±(0.4% of reading ± 0.2% of full scale)
High Accuracy: Option not available for units ranged under 5 scm or over 500 slpm
Repeatability: ±0.2% full scale
Zero Shift and Span Shift: 0.02% full scale/°Celsius/Atm
Operating Range/Turndown Ratio 0.5% to 100%
Full Scale/200: 1 turndown
Maximum Measurable Flow Rate: 128% full scale
Typical Response Time: 10 ms (adjustable)
Warm-Up Time: <1 second
Mass Reference Conditions (STP): 25°C (77°F) and 14.696 psia (standard-others available on request)
Operating Temperature: -10 to 50°C
Humidity Range: (Non-Condensing) 0 to 100%
Maximum Pressure: 145 psig [50 psig (-LP option)]
Ingress Protection: IP40
Wetted Materials: 303 and 302 stainless steel, FKM, Silicone RTV (Rubber), Glass Reinforced Nylon, Aluminum
Note: If your application demands a different material, please contact trodeks. Do not subject FMA-LP1600A differential pressure sensor to upstream-downstream pressure differentials exceeding 15 psid.
Programmed Gases: Acetylene, Air, Argon, Butane, Carbon Dioxide, Carbon Monoxide, Ethane, Ethylene, Helium, Hydrogen, Iso-Butane, Krypton, Methane, Neon, Nitrogen, Nitrous Oxide, Oxygen, Propane, Sulfur Hexafluoride, Xenon, A-25, A-75, A1025, C-2, C-8, C-10, C-25, C-75, P-5, Star29
LCD Display: Monochrome with integrated touchpad simultaneously displays mass flow, volumetric flow, pressure and temperature
Digital Output: RS232 and USB Serial
Electrical Supply: 4 AA batteries, 9 to 20 Vdc wall outlet adaptor minimum 150 mA
Note: The digital output signal communicates mass flow, volumetric flow, pressure and temperature.
Weight: 5.4 to 9.1 kg (12 to 20 lb), depending on configuration
Case Dimensions: 406 L x 330 W x 178 mm D (16 x 13 x 7")
Process Connections: 1/8, ¼, 3/8; or ½" push-connect style tubing fittings Anemometer | Automation | Accelerometers | Calibrator | Cartridge Heaters | Chart Recorders | Cryogenics | Custom Pressure Transducers | Data Loggers | Data Acquisition | Dial Thermometers | Fiber Optics | Fiber Optic Temperature Measurement | Flexible Heaters | Flow Meters | Glass Bulb Thermometers | Grippers | Heaters | Humidity Measurement | Integrated Circuit Sensors | Immersion Heaters | Infrared Thermometers | Level Measurement | Load Cells | M12 Temperature Connectors | Magnetic Flow Meter | Panel Meter | PID Controllers | pH meter | Pressure Gauges | Pressure Transducers | Pressure Transmitters | Rotameters | RTD | Remote Process Monitoring | Sanitary Sensors | Signal Conditioners | Stepper Motors | Strain Gages | Strip Heaters | Temperature Controllers | Temperature Labels | Temperature Measurement | Temperature Sensor Manufacturing | Temperature Sensors | Temperature Transmitters | Thermal Imagers | Thermistors | Thermocouple | Thermocouple Wire | Thermowell | Ultrasonic Flow Meters | Vortex Flow Meter | Wireless Sensors