Dissolved Oxygen Transmitter- Analyzer and Controller
DOTX441
PLACE ORDER - Analyzer/Transmitter/Controller
- NEMA 7X (IP68) Enclosure
- Galvanic Cell for PPB
- Self-Guided Menu
- Automatic Calibration and Check
- ON/OFF/PWM Control
- Programmable Transmitter Output
- RS485 Output, in MODBUS®
Description
Description
The DOTX441 dissolved oxygen analyzer issupplied in a compact cast aluminum case. SAE 323,with low oxidation grade, anticorrosion treatment andfinished with electrostatic epoxy paint. The DOTX441 iscompact, light weight, rugged and IP-68 rated. The units includelocal display, transmitter output and controller function allprogrammed via a 3-button side panel with an easy self guidedmenu. The DOTX441 can be mounted on a 51 mm (2") tube or flatsurface. Electrical connections are on the terminal bar located inthe back housing of the instrument with cable entrance throughtwo cable compression fittings of 13 mm (½") BSP. Commonapplications include food and drink, pulp and paper, chemical,sanitation, water and wastewater treatments.
SPECIFICATIONS
Case Cast: Aluminum SAE 323
Finishing: Electrostatic epoxy paint
Electrical Connection Connector Bar: Conexell type
Cable Inlet Cable Knockout: 13 mm (½") (2 x)
Assembly: 51 mm (2") tube or flat surface
Identification: SS tag
Enclosure: NEMA 7X (IP68)
Power Consumption: 3.5 VA
Weight: 1.9 kg (4.3 lb)
Electrical Power: 90 to 240 Vac (50/60 Hz)
Working Temperature: 5 to 40°C (41 to 104°F)
Relative Humidity: 20 to 80%
Analyzer/Transmitter
Display Alphanumeric: 2 lines x 16 characters
Range: 0 to 60 mg O2/L
Resolution: 0.1 or 0.01
Relative Precision: 0.02% (full scale)
Working Temperature: 50°C (122°F)
Auto Temperature Compensation: 50°C (122°F)
Measuring Principle: Clark polarographic cell
Response Time: 20 s (programmable)
Barometric Pressure Compensation: Up to 10.000 m
Signal Output Analogic: 4 to 20 mA with programmable output range for all reading range
Galvanic Isolation: 2000 Vac (by opto coupler)
Line Resistance: 1 k&trodeks;
Controller
Actuation Type Frequency: Modulation, P+D
Set-Points: 2 independent from 0 to 100% of scale
Outputs: 2 control ON/OFF, programmable for alarm or actuation P+D
Control for Automatic Cleaning: Timer ON/OFF, for periods up to 99 seconds, and in intervals up to 99 hours
Cell
Type: Immersion
Body material: polypropylene (PP) or 316 SS
Maximum Temperature: (@ 1 ATM) 50°C (122°F)
Body Length: 1.5 m (59"), standard
Auto Cleaning System: Activated by a solenoid valve Anemometer | Automation | Accelerometers | Calibrator | Cartridge Heaters | Chart Recorders | Cryogenics | Custom Pressure Transducers | Data Loggers | Data Acquisition | Dial Thermometers | Fiber Optics | Fiber Optic Temperature Measurement | Flexible Heaters | Flow Meters | Glass Bulb Thermometers | Grippers | Heaters | Humidity Measurement | Integrated Circuit Sensors | Immersion Heaters | Infrared Thermometers | Level Measurement | Load Cells | M12 Temperature Connectors | Magnetic Flow Meter | Panel Meter | PID Controllers | pH meter | Pressure Gauges | Pressure Transducers | Pressure Transmitters | Rotameters | RTD | Remote Process Monitoring | Sanitary Sensors | Signal Conditioners | Stepper Motors | Strain Gages | Strip Heaters | Temperature Controllers | Temperature Labels | Temperature Measurement | Temperature Sensor Manufacturing | Temperature Sensors | Temperature Transmitters | Thermal Imagers | Thermistors | Thermocouple | Thermocouple Wire | Thermowell | Ultrasonic Flow Meters | Vortex Flow Meter | Wireless Sensors