Your Source for Process Measurement & Control
UK
44-796-072-6366
Germany
49-178-631-1266
Analog Input Expansion Module for OMB-DAQ-3000 Series and OMB-DAQBOARD-3000 Series
OMB-PDQ30
PLACE ORDER - Adds 24 Differential/48 Single-Ended Inputs to OMB‑DAQBOARD‑3000 Series PCI Boards and OMB‑DAQ‑3000 Series Personal Daq USB Data Acquisition Modules
- Any Input Can Be a Voltage or a Thermocouple
- Convenient Removable Screw-Terminal Connections
Description
Description
The OMB-PDQ30 analog inputexpansion module increasestotal channel capacity of anOMB‑DAQBOARD‑3000Series PCI bus data acquisitionboard or OMB-DAQ-3000 SeriesPersonal Daq USB data acquisitionmodule to a total of 32 differentialor 64 single-ended analog inputs.The OMB‑PDQ30 attaches toan OMB-DAQBOARD-3000Series board via an optionalOMB-CA-266-3 cable, and to theOMB-DAQ-3000 Series either byplugging directly into the PersonalDaq module with included clips,or via an optional OMB-CA-96cable if it is necessary to havesome distance between devices.
Voltage measurements via theOMB-PDQ30 have the same rangesas the host OMB-DAQBOARD-3000Series PCI board or OMB-DAQ-3000Series Personal Daq USBmodule.When used with either host,the OMB-PDQ30 also addsthermocouple measurementcapability, in which the A/Doperates in a programmable,over-sample/digital filtering modeto provide low-noise and stabletemperature measurements.
Specifications
GENERAL
Operating Temperature:
-30 to 70°C (-22 to 158°F)
Storage Temperature:
-40 to 80°C (-40 to 176°F)
Warm-Up: 30 min to ratedspecifications
Relative Humidity:
0 to 95%, non-condensing
Vibration: MIL STD 810E,category 1 and 10
Communications Connector: 25-pin DSUB
Signal I/O Connector:
6 removable screw-terminalblocks
(12 connections each)
Dimensions:
269 W x 92 D x 45 mm H
(10.6 x 3.6 x 1.6")
Weight: 400 g (0.88 lbs)
Power: Supplied byOMB-DAQBOARD-3000 Series
PCIboard or OMB-DAQ-3000 Series
Personal Daq USB module;400 mW max
ANALOG INPUTS
Channels: 48 single-ended inputs,
24 channels of differential inputs,
Voltage Measurement Speed:
1 µs per channel
Temperature Measurement Speed:
Programmable from 100 µs to 20 msper channel
Ranges: ±10 V, ±5 V, ±2 V,
±1 V, ±500 mV, ±200 mV, ±100 mV,
universal thermocouple
Total Harmonic Distortion:
-80 dB typical for ±10 V range,
1 kHz fundamental
Signal to Noise and Distortion:
72 dB typical for ±10 V range, 1 kHz
Fundamental Input Impedance:
10 MΩ (single-ended); 20 MΩ (differential)
Bias Current:
40 pA typical (0 to 35°C)
Over-Voltage Protection: ±30 V
Ranges: Software or sequencer
selectableon a per-channel basis
Crosstalk: -90 dB DC to 60 Hz;
86 dB @10 kHz, typical Anemometer | Automation | Accelerometers | Calibrator | Cartridge Heaters | Chart Recorders | Cryogenics | Custom Pressure Transducers | Data Loggers | Data Acquisition | Dial Thermometers | Fiber Optics | Fiber Optic Temperature Measurement | Flexible Heaters | Flow Meters | Glass Bulb Thermometers | Grippers | Heaters | Humidity Measurement | Integrated Circuit Sensors | Immersion Heaters | Infrared Thermometers | Level Measurement | Load Cells | M12 Temperature Connectors | Magnetic Flow Meter | Panel Meter | PID Controllers | pH meter | Pressure Gauges | Pressure Transducers | Pressure Transmitters | Rotameters | RTD | Remote Process Monitoring | Sanitary Sensors | Signal Conditioners | Stepper Motors | Strain Gages | Strip Heaters | Temperature Controllers | Temperature Labels | Temperature Measurement | Temperature Sensor Manufacturing | Temperature Sensors | Temperature Transmitters | Thermal Imagers | Thermistors | Thermocouple | Thermocouple Wire | Thermowell | Ultrasonic Flow Meters | Vortex Flow Meter | Wireless Sensors