Differential Pressure and Thermocouple Data Logger
OM-CP-PRTC110
PLACE ORDER - 5 Year Battery Life
- Wide Temperature Range
- Accepts Thermocouple Types
J, K, T, E, R, S, B, and N - High Speed Download
- Real-Time Operation
- Programmable Start Time
- Miniature Size
Description
Description
The OM-CP-PRT110 is a differentialpressure and temperature datalogger, ideal for use in factory andlaboratory applications.The OM-CP-PRT110 features areal-time clock module that extendsthe battery life to >2 years andallows for high speed downloads.This device will measure and recordup to 10,922 measurements perchannel. The storage medium isnon-volatile solid state memory,providing maximum data securityand stopped directly from yourcomputer and its small size allows itto fit almost anywhere.The OM-CP-PRTC110 makes dataretrieval quick and easy. Simplyplug it into an available com portand our user-friendly software doesthe rest. The software converts a PCinto a real-time strip chart recorder.Data can be printed in graphical andtabular format or exported to a text orMicrosoft Excel® file.
Specifications
INTERNAL CHANNEL
Temperature Range: -20 to 80°C (-4 to 176°F)
Temperature Resolution: 0.1°C
Calibrated Accuracy: ±0.5°C (0 to 50°C)
REMOTE CHANNELS
Thermocouple Types: J, K, T, E, R, S, B, and N
Thermocouple Connection: Female subminiaturethermocouple connector
Cold Junction Compensation: Automatic, based on internalchannel
Maximum Thermocouple Resistance: 1000 Ω, <1000 Ω recommended
Pressure Range (Nominal): 0 to 2 inH2O or 0 to 5 inH2O(differential)
Pressure Range (Max): -2 to +4 inH2Oand -2 to +8inH2O (differential)
Pressure Connections: REF and IN ports via 1/16" ID, 5/32" Max OD,polyurethane tubing
Pressure Resolution: 0.002 inH2O and 0.005 inH2O
Pressure Specified Accuracy: Nominal range at 25°C
Combined Calibration Accuracy: ±0.5°C FSR maxat 0.25% over 1 year
Start Modes: Software programmable immediatestart or delay start up six months in advance
Real-Time Recording: May be used with PC to monitorand record data in real time
Memory: 10,922 readings per channel; software configurablememory wrap
Reading Rate: 2 seconds to every 24 hours, selectable in software
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type: 3.6V lithium battery (included), user replaceable
Battery Life: 2 years (15 minute reading rate at 25°C)
Data Format: Date and time stamped °C, °F, K, °R; inH2O,PSIG,inHg, mmHG, bar,
atm, Torr, Pa, kPa, and MPa
Time Accuracy: ±1 minute/month (at 20 to 30°C)
Computer Interface: PC serial or USB (interface cable required);57,600 baud
Software: XP SP3/Vista/7 and 8(32-bit and 64-bit)
Operating Environment:-20 to 80°C (-4 to 176°F), 0 to 95% RH non-condensing
Dimensions: 68 H x 41 W x 20 mm D (2.7 x 1.6 x 0.8") plusconnective tube
Enclosure: ABS plastic
Weight: 250 g (8.8 oz) Anemometer | Automation | Accelerometers | Calibrator | Cartridge Heaters | Chart Recorders | Cryogenics | Custom Pressure Transducers | Data Loggers | Data Acquisition | Dial Thermometers | Fiber Optics | Fiber Optic Temperature Measurement | Flexible Heaters | Flow Meters | Glass Bulb Thermometers | Grippers | Heaters | Humidity Measurement | Integrated Circuit Sensors | Immersion Heaters | Infrared Thermometers | Level Measurement | Load Cells | M12 Temperature Connectors | Magnetic Flow Meter | Panel Meter | PID Controllers | pH meter | Pressure Gauges | Pressure Transducers | Pressure Transmitters | Rotameters | RTD | Remote Process Monitoring | Sanitary Sensors | Signal Conditioners | Stepper Motors | Strain Gages | Strip Heaters | Temperature Controllers | Temperature Labels | Temperature Measurement | Temperature Sensor Manufacturing | Temperature Sensors | Temperature Transmitters | Thermal Imagers | Thermistors | Thermocouple | Thermocouple Wire | Thermowell | Ultrasonic Flow Meters | Vortex Flow Meter | Wireless Sensors